Artifacts: cause and fix

Aliasing (wrap-around) in MRI: cause and fixes

Anatomy outside the field of view folds back into the image. These are the fixes and what each one costs.

Aliasing (wrap-around)

Appearance

Anatomy outside the FOV appears folded onto the opposite side of the image.

Cause

Tissue outside the FOV in the phase direction is still excited and receives phase values identical to tissue inside the FOV. In 3D it also occurs in the slice direction.

How to fix
  • Phase oversampling (No Phase Wrap / Foldover suppression)
  • Enlarge the FOV
  • Swap phase to the shorter anatomical axis
  • Saturation band on the outside tissue
  • Use a coil that does not receive from the outside tissue; add slice oversampling in 3D
Cost of the fix

Oversampling adds time or reduces averages. A larger FOV lowers resolution unless the matrix is raised.

Phase oversampling (No Phase Wrap)

Purpose

Prevents wrap-around when anatomy extends beyond the FOV in the phase direction.

How it works

The scanner acquires a larger FOV in the phase direction and discards the extra part.

Consequences

Extra phase lines add time unless averages are reduced to compensate. GE No Phase Wrap doubles the phase FOV and halves NEX, so time and SNR stay about the same.

Practical tip

Use only as much oversampling as the anatomy needs. A saturation band over the outside tissue is an alternative when time is tight.

This page is one entry from the MRI Technologist Reference

The full reference has 31 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.