عربي
Protocols by body region

Brain MRI planning and protocol: slice positioning and routine sequences

How to plan a routine brain (head) MRI: slice planning drawings, plus one common protocol with what each sequence is there to show and when contrast is added.

Protocols are not universal. They differ from one hospital to another and from one radiologist to another, and they change with the scanner, the coil, the software version and the clinical question. What is written here is one common way of doing each examination that agrees with the published guidance listed in the references. It does not have to match your department, and it does not replace your department's approved protocol or the radiologist's request.

Brain: routine

Indications

Headache, non-specific neurological symptoms, first-line survey.

Set-up

Head coil, supine, head first. Axials parallel to the AC-PC line. FOV 22–24 cm, 5 mm slices, 1 mm gap.

Slice planning
ACPCAP
Axial Parallel to the AC-PC line, from the foramen magnum to above the vertex.
RL
Axial, checked on the coronal Tilt the slices until the two sides are symmetric.
RLAP
Sagittal Parallel to the midline (falx), from one temporal lobe to the other.

Schematic drawing, not a patient image.

Sequences and why
Sag T1Midline anatomy: corpus callosum, pituitary, craniocervical junction. Also used to plan the axials.
Ax T2 FSEEdema, CSF spaces, vascular flow voids, posterior fossa.
Ax T2 FLAIRCSF is nulled, so periventricular and cortical lesions stand out.
Ax DWI b0 / b1000 + ADCAcute infarct, abscess, highly cellular tumour.
Ax SWI or T2* GREBlood products, microbleeds, calcification.
Cor T2 or Cor FLAIRTemporal lobes, hippocampi and a second plane to confirm lesions.
Practical points
  • Copy the same slice positions to every axial sequence so lesions can be compared slice by slice
  • Axial phase direction right-left keeps eye-movement ghosts off the brain; DWI uses anterior-posterior
  • Check that the top slice covers the vertex and the bottom slice covers the foramen magnum

This page is one entry from the MRI Technologist Reference

The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.