Shoulder MRI planning and protocol: slice positioning and sequences
How to plan a shoulder MRI: slice planning drawings for the oblique planes, plus one common protocol with what each sequence shows, and how MR arthrography differs.
Protocols are not universal. They differ from one hospital to another and from one radiologist to another, and they change with the scanner, the coil, the software version and the clinical question. What is written here is one common way of doing each examination that agrees with the published guidance listed in the references. It does not have to match your department, and it does not replace your department's approved protocol or the radiologist's request.
Shoulder
Rotator cuff tear, impingement, instability, labral tear.
Shoulder coil, supine, arm at the side in neutral or slight external rotation, weight on the forearm. FOV 14–16 cm, 3 mm slices.
Schematic drawing, not a patient image.
| Ax PD fat-sat | Labrum, subscapularis, biceps tendon in its groove. |
| Cor oblique PD or T2 fat-sat | Supraspinatus tear, superior labrum, AC joint. Planned along the supraspinatus tendon. |
| Cor oblique T1 | Marrow, anatomy, fat. |
| Sag oblique T2 fat-sat | Cuff tendons in cross-section, rotator interval. |
| Sag oblique T1 extended medially | Fatty atrophy of the cuff muscles, which decides whether a tear is repairable. |
- Internal rotation overlaps the cuff tendons and hides tears
- Magic angle brightens the supraspinatus on short TE: confirm on T2
- The shoulder is off-centre, so spectral fat sat is often uneven: use Dixon or STIR
- Breathing ghosts come from the chest wall: choose the phase direction so they miss the joint
This page is one entry from the MRI Technologist Reference
The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.
Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.