Flip angle in MRI: what it does to contrast and SAR
The flip angle controls T1 weighting in gradient echo and drives SAR in spin echo. This is how to choose it.
Flip angle
In GRE it is the main control of T1 weighting and signal. In FSE the refocusing angle controls SAR.
The angle by which the RF pulse tips magnetization away from the z-axis. For a given TR and T1 the signal peaks at the Ernst angle: cos(angle) = exp(-TR/T1).
Spoiled GRE: larger angle gives more T1 weighting; small angle (5–20°) gives PD or T2* weighting. SAR rises with the square of the flip angle, so lowering FSE refocusing from 180° to 130° roughly halves SAR with a small SNR cost.
When the scanner stops for SAR, reduce the refocusing flip angle before you cut slices or lengthen TR.
This page is one entry from the MRI Technologist Reference
The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.
Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.