Parameters and their trade-offs

NEX, NSA and averages in MRI: what doubling really buys

Doubling the averages doubles the time and gives only 41% more SNR. This is when it is worth it.

NEX / NSA / Averages

Purpose

Raises SNR by repeating the acquisition and averaging.

How it works

Signal adds coherently, noise adds randomly: SNR grows with √NEX while time grows with NEX.

Consequences

Going from 1 to 2 NEX doubles the time for a 41% SNR gain. Going from 2 to 4 doubles it again for another 41%. Averaging also smooths motion and flow ghosts.

Practical tip

This is the most expensive way to buy SNR. Thicker slices, a slightly larger FOV, narrower bandwidth or deep-learning reconstruction are cheaper.

This page is one entry from the MRI Technologist Reference

The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.