MRCP planning and protocol: slab positioning and sequences

How to plan an MRCP: planning drawings for the slabs, plus one common protocol with what each sequence shows and when a dynamic contrast study is added.

Protocols are not universal. They differ from one hospital to another and from one radiologist to another, and they change with the scanner, the coil, the software version and the clinical question. What is written here is one common way of doing each examination that agrees with the published guidance listed in the references. It does not have to match your department, and it does not replace your department's approved protocol or the radiologist's request.

MRCP

Indications

Bile duct stones, biliary obstruction, strictures, pancreatic duct, anatomical variants before surgery.

Set-up

Torso array, respiratory bellows. Fasting 4–6 hours. A manganese-rich juice (pineapple, blueberry) can be given to darken stomach fluid.

Slice planning
CBDRLAP
Thick-slab MRCP Four to six radial slabs rotated around the common bile duct, planned on the axial image.
RL
3D MRCP Coronal volume covering the gallbladder, the bile ducts and the head of the pancreas.

Schematic drawing, not a patient image.

Sequences and why
Cor and Ax single-shot T2Anatomy overview; stones appear as dark filling defects.
Ax T2 fat-satLiver, pancreas, duct wall, fluid collections.
Thick-slab 2D single-shot MRCP, radial slabsWhole biliary tree in a 2-second breath-hold; TE near 700–1000 ms leaves only static fluid bright.
3D respiratory-triggered MRCP, about 1 mmSmall stones, strictures and variants on thin source images; MIP for overview.
Ax T1 fat-sat GREPancreatic parenchyma (normally bright), haemorrhage, protein plugs.
Practical points
  • Small stones hide in the MIP: read and send the source images
  • Coach steady, regular breathing for the triggered 3D scan; if breathing is irregular use a breath-hold 3D or the thick slabs
  • Pulsation of the hepatic artery across the duct mimics a stone or stricture
  • Air in the ducts floats to the non-dependent side; stones sink

This page is one entry from the MRI Technologist Reference

The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.