Fat suppression methods in MRI: fat sat, SPAIR, STIR and Dixon
Four ways to make fat dark. Each one fails in a different situation, so the choice depends on the anatomy.
Fat suppression method
Makes edema, enhancement and fluid visible next to fat, and confirms that a lesion contains fat.
Spectral fat sat (CHESS): frequency-selective pulse, fast, needs a homogeneous field. SPAIR: adiabatic inversion, tolerates B1 variation, slower. STIR: nulls by T1, works in poor homogeneity and at low field, lower SNR. Dixon: calculated from phase, most uniform, longer scan. Water excitation: excites water only, good for cartilage and 3D GRE.
Spectral methods fail off-centre, near metal, air and curved anatomy (neck, shoulder, ankle, breast edge), and can suppress water instead of fat. STIR is not specific to fat. Dixon can swap fat and water in isolated regions.
Rule of thumb: large or off-centre FOV and metal: STIR or Dixon. Post-contrast: spectral, SPAIR or Dixon, never STIR. Place the shim volume on the anatomy, not on air.
This page is one entry from the MRI Technologist Reference
The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.
Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.