TR (repetition time) in MRI: what it changes
TR sets T1 weighting and most of the scan time. This is what happens to the image and the clock when you move it.
TR (Repetition Time)
Sets how much longitudinal magnetization recovers between excitations, which controls T1 weighting.
Time from one excitation pulse to the next on the same slice. Tissues with short T1 (fat, gadolinium-enhanced tissue) recover fully within a short TR; tissues with long T1 (CSF) do not.
Longer TR: more SNR, less T1 weighting, more slices per TR, longer scan (time is proportional to TR). Shorter TR: more T1 weighting, less SNR, fewer slices, and higher SAR per unit time.
If the slices do not fit in a T1 sequence, add a second acquisition (concatenation) instead of raising TR beyond about 700 ms, or you lose T1 contrast.
This page is one entry from the MRI Technologist Reference
The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.
Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.