Lumbar spine MRI planning and protocol: slice positioning and sequences

How to plan a lumbar spine (lower back) MRI: slice planning drawings, plus one common protocol with what each sequence shows and when contrast is added.

Protocols are not universal. They differ from one hospital to another and from one radiologist to another, and they change with the scanner, the coil, the software version and the clinical question. What is written here is one common way of doing each examination that agrees with the published guidance listed in the references. It does not have to match your department, and it does not replace your department's approved protocol or the radiologist's request.

Lumbar spine

Indications

Low back pain with sciatica, stenosis, cauda equina syndrome, post-operative assessment.

Set-up

Spine coil, cushion under the knees. Landmark about 5 cm above the iliac crest. Sagittal 4 mm, FOV 28–30 cm, T12 to S2.

Slice planning
L1L2L3L4L5S1RL
Sagittal Parallel to the spine on the coronal localizer, wide enough to include both foramina.
T12L1L2L3L4L5S1AP
Axial Parallel to each disc, at least L3-4, L4-5 and L5-S1. Saturation band over the abdomen.

Schematic drawing, not a patient image.

Sequences and why
Sag T2 FSEDisc hydration and herniation, canal, conus.
Sag T1Marrow, endplate (Modic) changes, fat.
Sag STIREdema: acute fracture, infection, pars stress, active endplate change.
Ax T2 FSE parallel to the discsHerniation type, lateral recess, foramina, facet joints.
Ax T1Nerve roots outlined by foraminal fat.
Practical points
  • Anterior saturation band over the aorta and bowel
  • Always include the conus
  • A coronal STIR of the sacroiliac joints and hips adds about two minutes and often explains the pain

This page is one entry from the MRI Technologist Reference

The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.