Brain MRI planning and protocol: slice positioning and routine sequences
How to plan a routine brain (head) MRI: slice planning drawings, plus one common protocol with what each sequence is there to show and when contrast is added.
Protocols are not universal. They differ from one hospital to another and from one radiologist to another, and they change with the scanner, the coil, the software version and the clinical question. What is written here is one common way of doing each examination that agrees with the published guidance listed in the references. It does not have to match your department, and it does not replace your department's approved protocol or the radiologist's request.
Brain: routine
Headache, non-specific neurological symptoms, first-line survey.
Head coil, supine, head first. Axials parallel to the AC-PC line. FOV 22–24 cm, 5 mm slices, 1 mm gap.
Schematic drawing, not a patient image.
| Sag T1 | Midline anatomy: corpus callosum, pituitary, craniocervical junction. Also used to plan the axials. |
| Ax T2 FSE | Edema, CSF spaces, vascular flow voids, posterior fossa. |
| Ax T2 FLAIR | CSF is nulled, so periventricular and cortical lesions stand out. |
| Ax DWI b0 / b1000 + ADC | Acute infarct, abscess, highly cellular tumour. |
| Ax SWI or T2* GRE | Blood products, microbleeds, calcification. |
| Cor T2 or Cor FLAIR | Temporal lobes, hippocampi and a second plane to confirm lesions. |
- Copy the same slice positions to every axial sequence so lesions can be compared slice by slice
- Axial phase direction right-left keeps eye-movement ghosts off the brain; DWI uses anterior-posterior
- Check that the top slice covers the vertex and the bottom slice covers the foramen magnum
This page is one entry from the MRI Technologist Reference
The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.
Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.