Shoulder MRI planning and protocol: slice positioning and sequences

How to plan a shoulder MRI: slice planning drawings for the oblique planes, plus one common protocol with what each sequence shows, and how MR arthrography differs.

Protocols are not universal. They differ from one hospital to another and from one radiologist to another, and they change with the scanner, the coil, the software version and the clinical question. What is written here is one common way of doing each examination that agrees with the published guidance listed in the references. It does not have to match your department, and it does not replace your department's approved protocol or the radiologist's request.

Shoulder

Indications

Rotator cuff tear, impingement, instability, labral tear.

Set-up

Shoulder coil, supine, arm at the side in neutral or slight external rotation, weight on the forearm. FOV 14–16 cm, 3 mm slices.

Slice planning
RLAP
Coronal oblique Parallel to the supraspinatus tendon, that is perpendicular to the glenoid.
RLAP
Sagittal oblique Parallel to the glenoid, from the scapular neck to the outer edge of the humeral head.
RL
Axial From above the acromioclavicular joint to below the glenoid.

Schematic drawing, not a patient image.

Sequences and why
Ax PD fat-satLabrum, subscapularis, biceps tendon in its groove.
Cor oblique PD or T2 fat-satSupraspinatus tear, superior labrum, AC joint. Planned along the supraspinatus tendon.
Cor oblique T1Marrow, anatomy, fat.
Sag oblique T2 fat-satCuff tendons in cross-section, rotator interval.
Sag oblique T1 extended mediallyFatty atrophy of the cuff muscles, which decides whether a tear is repairable.
Practical points
  • Internal rotation overlaps the cuff tendons and hides tears
  • Magic angle brightens the supraspinatus on short TE: confirm on T2
  • The shoulder is off-centre, so spectral fat sat is often uneven: use Dixon or STIR
  • Breathing ghosts come from the chest wall: choose the phase direction so they miss the joint

This page is one entry from the MRI Technologist Reference

The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.