Ankle MRI planning and protocol: slice positioning and sequences

How to plan an ankle MRI: slice planning drawings, plus one common protocol with what each sequence shows and when contrast is added.

Protocols are not universal. They differ from one hospital to another and from one radiologist to another, and they change with the scanner, the coil, the software version and the clinical question. What is written here is one common way of doing each examination that agrees with the published guidance listed in the references. It does not have to match your department, and it does not replace your department's approved protocol or the radiologist's request.

Ankle

Indications

Ligament sprain, tendon pathology, osteochondral lesion, Achilles tear.

Set-up

Ankle or knee coil, foot near 90°. FOV 14–16 cm, 3 mm slices.

Slice planning
AP
Axial Parallel to the tibiotalar joint, from above the joint to the sole.
AP
Coronal Perpendicular to the tibiotalar joint, from the back of the calcaneus to the midfoot. The sagittal slices are planned on the coronal, parallel to the long axis of the tibia and the calcaneus.

Schematic drawing, not a patient image.

Sequences and why
Sag T1Marrow, Achilles tendon, sinus tarsi fat.
Sag STIRMarrow edema, Achilles, plantar fascia.
Ax PD fat-satTendons in cross-section; anterior talofibular and calcaneofibular ligaments.
Ax T1 or PDTendon anatomy without fat suppression.
Cor PD fat-satTalar dome, deltoid and spring ligaments.
Practical points
  • Tendons curve around the malleoli at the magic angle: confirm bright signal on a long-TE sequence
  • The curved anatomy makes spectral fat sat unreliable: STIR or Dixon is safer
  • Angle the axials perpendicular to the tendons of interest

This page is one entry from the MRI Technologist Reference

The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.