Parameters and their trade-offs

b-value in DWI: how to choose it

A higher b-value gives more diffusion weighting and less signal. The right value depends on the organ.

b-value (DWI)

Purpose

Sets how strongly the image is weighted by water diffusion.

How it works

Two matched gradient lobes dephase and rephase spins. Water that moved between them is not fully rephased and loses signal. At least two b-values are needed to calculate the ADC map.

Consequences

Higher b: more diffusion contrast and less T2 shine-through, but lower SNR, longer TE and more distortion. Typical: brain 1000, liver 50 / 400 / 800, prostate 50–100, 800–1000 and a high b of 1400 or more.

Practical tip

Bright on DWI means restricted diffusion only when the ADC map is dark. Always send the ADC map.

This page is one entry from the MRI Technologist Reference

The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.