Cervical spine MRI planning and protocol: slice positioning and sequences

How to plan a cervical spine (neck) MRI: slice planning drawings, plus one common protocol with what each sequence shows and when contrast is added.

Protocols are not universal. They differ from one hospital to another and from one radiologist to another, and they change with the scanner, the coil, the software version and the clinical question. What is written here is one common way of doing each examination that agrees with the published guidance listed in the references. It does not have to match your department, and it does not replace your department's approved protocol or the radiologist's request.

Cervical spine

Indications

Neck pain with radiculopathy, myelopathy, trauma, demyelination.

Set-up

Neck or spine coil, landmark at C4. Sagittal 3 mm, FOV 24 cm; axial 3–4 mm parallel to the discs.

Slice planning
C2C3C4C5C6C7T1RL
Sagittal Parallel to the spine on the coronal localizer, wide enough to include both foramina.
C2C7T1AP
Axial Parallel to each disc from C2-3 to C7-T1, or one block for cord disease. Saturation band over the pharynx.

Schematic drawing, not a patient image.

Sequences and why
Sag T2 FSECord signal, disc herniation, canal diameter.
Sag T1Marrow, alignment, fracture.
Sag STIRMarrow edema, ligament injury, cord lesions.
Ax T2* multi-echo GREForamina and disc versus osteophyte, with less CSF flow artifact than FSE.
Ax T2 FSECord lesions in cross-section.
Practical points
  • Anterior saturation band over the pharynx; ask the patient not to swallow during acquisition
  • GRE exaggerates foraminal narrowing because of susceptibility
  • CSF pulsation is strong at this level: use flow compensation

This page is one entry from the MRI Technologist Reference

The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.