Parameters and their trade-offs

ETL and turbo factor in MRI: speed against blurring

A longer echo train shortens the scan and blurs the image. This is where the limit is.

ETL / Turbo factor

Purpose

Shortens FSE scan time by collecting several phase-encoding lines after each excitation.

How it works

Scan time is divided by the ETL. Each echo in the train has a different TE; the one at the centre of k-space sets the contrast and the others shape the edges.

Consequences

Longer ETL: faster, but more blurring (worst with short effective TE), brighter fat (J-coupling), more SAR and fewer slices per TR. Short TE with long ETL is the classic cause of a blurred PD image.

Practical tip

Guide values: T1 FSE 2–4, PD 6–10, T2 12–20. Raising receiver bandwidth shortens echo spacing and reduces blurring at the same ETL.

This page is one entry from the MRI Technologist Reference

The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.