Fat and water frequency difference in MRI

Fat and water resonate at slightly different frequencies. Fat sat, chemical shift and Dixon all come from this one fact.

Fat and water are slightly out of tune

The idea

Protons in fat spin a little slower than protons in water: about 220 Hz at 1.5 T. This one fact explains four things. Fat is drawn a few pixels away from its true place (chemical shift artifact). A pulse tuned to the fat frequency can remove fat (spectral fat saturation). Fat and water drift in and out of step every 2.2 ms, so at some echo times they cancel in a shared voxel (opposed-phase). And from two or three echoes the scanner can calculate separate fat and water images (Dixon).

At the console

Everything doubles at 3 T: the shift is 440 Hz and the in-phase and opposed-phase echo times are halved. Where the field is uneven, fat is no longer at the expected frequency and spectral fat suppression fails.

This page is one entry from the MRI Technologist Reference

The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.