Parameters and their trade-offs

FOV in MRI: resolution, SNR and wrap

A smaller FOV gives finer detail, less signal and a risk of aliasing. This is how to balance it.

FOV (Field of View)

Purpose

Sets anatomical coverage and, with the matrix, the pixel size.

How it works

Pixel size = FOV ÷ matrix in each direction.

Consequences

SNR follows the square of the FOV: going from 24 to 18 cm leaves 56% of the SNR. A FOV smaller than the anatomy in the phase direction causes wrap-around.

Practical tip

Use a rectangular FOV with the short axis of the anatomy in the phase direction: fewer phase lines, same resolution, less time.

This page is one entry from the MRI Technologist Reference

The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.