Inversion recovery explained: how STIR and FLAIR switch a tissue off
An inversion pulse and a waiting time remove the signal of one tissue. That is all STIR and FLAIR are.
Inversion recovery: switching one tissue off
A 180° pulse turns all magnetization upside down. Each tissue then climbs back at its own T1 speed and passes through zero on the way. If you take the picture exactly when one tissue is at zero, that tissue gives no signal. Fat passes zero early, CSF much later.
Short TI (about 150 ms at 1.5 T) switches off fat: STIR. Long TI (about 2200–2500 ms) switches off CSF: FLAIR. The right TI changes with field strength, and in the heart it changes minute by minute after contrast.
This page is one entry from the MRI Technologist Reference
The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.
Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.