Parameters and their trade-offs

Parallel imaging factor (ASSET, SENSE, GRAPPA): how far to push it

Acceleration shortens the scan and costs SNR. The coil decides how high you can go.

Parallel imaging factor (R)

Purpose

Cuts scan time, or shortens the EPI echo train to reduce distortion.

How it works

Phase lines are skipped and the missing data is rebuilt from the different sensitivity profiles of the coil elements.

Consequences

SNR falls by at least √R, plus a g-factor penalty that shows as noise bands in the centre of the image. Too high a factor, or a FOV smaller than the anatomy with SENSE-type methods, puts a fold-over ghost in the middle of the image.

Practical tip

Acceleration only works along a direction with several coil elements. Keep R at 2 or below for 2D unless the coil has many channels; 3D can accelerate in two directions.

This page is one entry from the MRI Technologist Reference

The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.