Parameters and their trade-offs

Partial Fourier (half scan) in MRI

Part of k-space is skipped and filled in by calculation. It saves time or shortens TE, at a cost.

Partial Fourier

Purpose

Saves time (phase direction) or shortens TE (frequency direction).

How it works

k-space is symmetric, so a little more than half is acquired and the rest is calculated.

Consequences

Fewer measured lines mean less SNR and more sensitivity to motion and phase errors. Resolution is nominally kept, with slight blurring.

Practical tip

It is what makes SSFSE / HASTE possible. Avoid combining aggressive partial Fourier with high parallel imaging on a low-SNR sequence.

This page is one entry from the MRI Technologist Reference

The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.