عربي
Parameters and their trade-offs

Phase-encoding direction in MRI: how to choose it

The phase direction decides where motion ghosts and wrap appear. Choose it so they miss the anatomy you need.

Phase-encoding direction

Purpose

Decides where motion ghosts, flow ghosts and wrap-around appear in the image.

How it works

Motion and pulsation artifacts always spread along the phase direction, whatever the direction of the motion. Chemical shift and susceptibility distortion appear along the frequency direction (along phase in EPI).

Consequences

The wrong choice throws ghosts from the aorta, eyes, heart or bowel across the area of interest. Putting phase along the long axis of the anatomy needs phase oversampling and costs time.

Practical tip

Before each scan ask two questions: what moves here, and where will its ghost land? Then swap, or add a saturation band over the moving structure.

This page is one entry from the MRI Technologist Reference

The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.