Receiver bandwidth in MRI: SNR against chemical shift and metal
A narrow bandwidth gives more signal and more chemical shift. A wide one does the opposite and helps near metal.
Receiver bandwidth
Trades SNR against chemical shift, susceptibility distortion and minimum TE.
The range of frequencies sampled across the FOV during readout. Wider bandwidth means a steeper readout gradient and a shorter sampling time.
Doubling bandwidth costs about 30% SNR (SNR is proportional to 1/√BW), but halves chemical-shift misregistration and metal distortion, shortens minimum TE and echo spacing. Narrow bandwidth does the opposite.
Near metal use the widest bandwidth available and recover SNR with averages. At 3 T chemical shift doubles, so use about twice the 1.5 T bandwidth.
This page is one entry from the MRI Technologist Reference
The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.
Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.