Saturation bands in MRI: where to put them
A sat band removes signal from tissue that would send ghosts into the image. This is where it helps and what it costs.
Saturation bands
Remove signal from tissue that would send motion, flow or wrap artifacts into the image.
A 90° pulse followed by a spoiler gradient destroys magnetization in a slab just before each excitation.
Each band adds RF (SAR) and time inside the TR, so fewer slices fit. A band above or below the slices also removes inflowing blood signal: that selects arteries or veins in TOF.
Typical uses: anterior band for spine (swallowing, aorta, bowel), band over the abdominal wall in pelvis, superior band for arterial TOF of the head.
This page is one entry from the MRI Technologist Reference
The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.
Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.