Slice thickness and gap in MRI
Thin slices show small structures and lose signal. The gap controls cross-talk and coverage.
Slice thickness and gap
Sets through-plane resolution, coverage and partial-volume effect.
SNR is directly proportional to slice thickness. The gap keeps imperfect slice profiles from exciting their neighbours.
Thinner slices: less partial volume and better small-lesion detection, but less SNR and more slices to cover the anatomy. No gap: cross-talk lowers SNR and alters contrast unless slices are interleaved.
A gap of 10–20% of slice thickness is usual for 2D FSE. When you need contiguous slices, interleave or go 3D.
This page is one entry from the MRI Technologist Reference
The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.
Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.