TE (echo time) in MRI: what it changes

TE sets T2 weighting. A longer TE gives more T2 contrast and less signal.

TE (Echo Time)

Purpose

Sets how much transverse magnetization has decayed when the echo is read, which controls T2 (or T2*) weighting.

How it works

Time from excitation to the centre of the echo. In FSE the value that matters is the effective TE: the echo placed at the centre of k-space.

Consequences

Longer TE: more T2 weighting, fluid stands out, but SNR falls and susceptibility and flow voids grow. Shorter TE: more SNR, but the magic-angle effect appears in tendons on short-TE sequences (below about 32 ms).

Practical tip

In GRE at 1.5 T fat and water are opposed-phase at about 2.2 ms and in-phase at about 4.4 ms (about 1.1 and 2.2 ms at 3 T). Choose TE deliberately.

This page is one entry from the MRI Technologist Reference

The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.