Parameters and their trade-offs

TI (inversion time) in MRI: STIR, FLAIR and nulling a tissue

TI decides which tissue is switched off. The right value depends on the tissue and on the field strength.

TI (Inversion Time)

Purpose

Nulls the signal of one tissue (fat in STIR, CSF in FLAIR, normal myocardium in LGE) or increases T1 contrast.

How it works

A 180° pulse inverts magnetization. Each tissue passes through zero at TI = 0.69 × its T1. Exciting at that moment produces no signal from that tissue.

Consequences

A wrong TI leaves the target tissue partly bright. T1 values rise with field strength, so a TI copied from 1.5 T fails at 3 T. STIR also suppresses anything with fat-like T1, including gadolinium-enhanced tissue.

Practical tip

Never use STIR as the post-contrast fat-suppression method. Use spectral fat sat or Dixon.

This page is one entry from the MRI Technologist Reference

The full reference has 32 protocols, 9 dynamic study timelines, 19 artifacts, 16 scanner problems and 15 safety topics, as a PDF in English and Arabic plus an interactive tool.

Educational reference. It does not replace your site protocols, the radiologist's instructions, the MR safety officer, or the manufacturer's instructions for any implant, device or contrast agent. Parameter values are typical starting points at 1.5 T and must be adapted to the scanner, coil and patient.